Recall Z-2063-2018— Orthofix Srl

Class II · Terminated

Class II FDA medical device recall by Orthofix Srl, initiated 2018-01-30, terminated 2021-05-13. It names one FDA 510(k) clearance: K170650. Product: TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions.. Reason: The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts..

Recalling firm
Orthofix Srl
Classification
Class II
Status
Terminated
Initiated
2018-01-30
Terminated
2021-05-13
Firm location
Bussolengo, N/A, Italy

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions.

Reason for recall

The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.

Data sourced from openFDA. This site is unofficial and independent of the FDA.