Recall Z-2062-2020— Innovasis, Inc

Class II · Terminated

Class II FDA medical device recall by Innovasis, Inc, initiated 2020-04-07, terminated 2022-01-26. It names one FDA 510(k) clearance: K192049. Product: Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical Plate System is intended for use in anterior cervical fixation.. Reason: Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw..

Recalling firm
Innovasis, Inc
Classification
Class II
Status
Terminated
Initiated
2020-04-07
Terminated
2022-01-26
Firm location
Salt Lake City, UT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical Plate System is intended for use in anterior cervical fixation.

Reason for recall

Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.