Recall Z-2059-2014— Mindray DS USA, Inc. dba Mindray North America

Class II · Terminated

Class II FDA medical device recall by Mindray DS USA, Inc. dba Mindray North America, initiated 2014-05-19, terminated 2017-04-20. FDA's recall record names no specific 510(k) clearance for this event. Product: V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.. Reason: ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect..

Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Classification
Class II
Status
Terminated
Initiated
2014-05-19
Terminated
2017-04-20
Firm location
Mahwah, NJ, United States

Product description

V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

Reason for recall

ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect.

Data sourced from openFDA. This site is unofficial and independent of the FDA.