Recall Z-2057-2013— Synthes (USA) Products LLC
Class II · Terminated
Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2013-05-31, terminated 2015-09-21. It names 3 FDA 510(k) clearances: K092556, K083694, K071184. Product: Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.. Reason: The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm.
- Recalling firm
- Synthes (USA) Products LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-05-31
- Terminated
- 2015-09-21
- Firm location
- West Chester, PA, United States
510(k) clearances named in this recall
3 on FDA's record- K092556 — SYNTHES (USA) 2.4MM VA-LCP TWO-COLUMN NARROW VOLAR DISTAL RADIUS
- K083694 — SYNTHES (USA) 2.4 MM VA-LCP TWO COLUMN VOLAR DISTAL RADIUS PLATES
- K071184 — SYNTHES (USA) VARIABLE ANGLE-LOCKING COMPRESSION PLATE (VA-LCP) DISTAL RADIUS SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Reason for recall
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm