Recall Z-2056-2012— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2012-06-22, terminated 2013-03-06. It names one FDA 510(k) clearance: K022947. Product: System 1 Base 220/240V The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.. Reason: Terumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System 1 centrifugal system triggers error messages and alarms; therefore it is easily recognizable by the user. The potential hazards are dependent on the application.
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-06-22
- Terminated
- 2013-03-06
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
System 1 Base 220/240V The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System 1 centrifugal system triggers error messages and alarms; therefore it is easily recognizable by the user. The potential hazards are dependent on the application