Recall Z-2045-2021— Canon Medical System, USA, INC.
Class II · Ongoing
Class II FDA medical device recall by Canon Medical System, USA, INC., initiated 2021-05-21. It names one FDA 510(k) clearance: K170412. Product: Canon Vantage Tiitan 1.5T, Premium Open-Bore MRI System, REF: MRT-1510 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.. Reason: The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire..
- Recalling firm
- Canon Medical System, USA, INC.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-05-21
- Firm location
- Tustin, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Canon Vantage Tiitan 1.5T, Premium Open-Bore MRI System, REF: MRT-1510 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
Reason for recall
The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.
Data sourced from openFDA. This site is unofficial and independent of the FDA.