Recall Z-2043-2013— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2013-06-07, terminated 2014-12-15. FDA's recall record names no specific 510(k) clearance for this event. Product: 00597103110 Provisional CR ART SURF PROV 34/STR YEL 10, Rx, Sterile; 00597103112 Provisional CR ART SURF PROV 34/STR YEL 12, Rx, Sterile; 00597103114 Provisional CR ART SURF PROV 34/STR YEL 14, Rx, Sterile; 00597103117 Provisional CR ART SURF PROV 34/STR YEL 17, Rx, Sterile; 00597103120 Provisional CR ART SURF PROV 34/STR YEL 20, Rx, Sterile; 00597103123 Provisional CR ART SURF PROV 34/STR YEL 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.. Reason: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d.

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-06-07
Terminated
2014-12-15
Firm location
Warsaw, IN, United States

Product description

00597103110 Provisional CR ART SURF PROV 34/STR YEL 10, Rx, Sterile; 00597103112 Provisional CR ART SURF PROV 34/STR YEL 12, Rx, Sterile; 00597103114 Provisional CR ART SURF PROV 34/STR YEL 14, Rx, Sterile; 00597103117 Provisional CR ART SURF PROV 34/STR YEL 17, Rx, Sterile; 00597103120 Provisional CR ART SURF PROV 34/STR YEL 20, Rx, Sterile; 00597103123 Provisional CR ART SURF PROV 34/STR YEL 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason for recall

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Data sourced from openFDA. This site is unofficial and independent of the FDA.