Recall Z-2042-2018— NordicNeuroLab AS

Class II · Terminated

Class II FDA medical device recall by NordicNeuroLab AS, initiated 2014-12-15, terminated 2021-03-24. It names one FDA 510(k) clearance: K123306. Product: nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.. Reason: An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images..

Recalling firm
NordicNeuroLab AS
Classification
Class II
Status
Terminated
Initiated
2014-12-15
Terminated
2021-03-24
Firm location
Bergen, N/A, Norway

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.

Reason for recall

An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.