Recall Z-2039-2021— BioFire Diagnostics, LLC

Class II · Terminated

Class II FDA medical device recall by BioFire Diagnostics, LLC, initiated 2021-06-07, terminated 2023-07-31. It names one FDA 510(k) clearance: K180966. Product: FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit). Reason: Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame..

Recalling firm
BioFire Diagnostics, LLC
Classification
Class II
Status
Terminated
Initiated
2021-06-07
Terminated
2023-07-31
Firm location
Salt Lake City, UT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Reason for recall

Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.

Data sourced from openFDA. This site is unofficial and independent of the FDA.