Recall Z-2038-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2026-03-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2026-03-19
Firm location
Northfield, IL, United States

Product description

Medline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.

Reason for recall

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Data sourced from openFDA. This site is unofficial and independent of the FDA.