Recall Z-2036-2023— Medtronic Neuromodulation

Class II · Ongoing

Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2023-06-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Vanta with AdaptiveStim Implantable Neurostimulator, Model 977006; spinal chord stimulation. Reason: A cardioversion procedure may damage the electronics in the Vanta INS device, making the INS unresponsive and non-functional. As of April 19, 2023, Medtronic has received a total of two complaints concerning this issue from patients implanted with a Vanta INS, both of which have resulted in explant procedures..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Ongoing
Initiated
2023-06-06
Firm location
Minneapolis, MN, United States

Product description

Vanta with AdaptiveStim Implantable Neurostimulator, Model 977006; spinal chord stimulation

Reason for recall

A cardioversion procedure may damage the electronics in the Vanta INS device, making the INS unresponsive and non-functional. As of April 19, 2023, Medtronic has received a total of two complaints concerning this issue from patients implanted with a Vanta INS, both of which have resulted in explant procedures.