Recall Z-2036-2012— Teleflex Medical
Class II · Terminated
Class II FDA medical device recall by Teleflex Medical, initiated 2012-06-12, terminated 2014-01-28. It names one FDA 510(k) clearance: K063758. Product: Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.. Reason: Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment..
- Recalling firm
- Teleflex Medical
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-06-12
- Terminated
- 2014-01-28
- Firm location
- Research Triangle Park, NC, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.
Reason for recall
Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.