Recall Z-2035-2012— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2012-07-02, terminated 2013-12-24. It names 4 FDA 510(k) clearances: K060937, K033357, K033326 and 1 more. Product: Philips Brilliance 6, 16, 40, 64, Big Bore, iCT and iCT SP are Computed Tomography X-Ray Systems. The devices are Computed Tomography Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. Reason: Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off..
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-07-02
- Terminated
- 2013-12-24
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
4 on FDA's record- K060937 — BRILLIANCE VOLUME
- K033357 — ACQSIM-MULTISLICE-CT, MODEL 453567042271
- K033326 — PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
- K012009 — MX8000 IDT CT SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Brilliance 6, 16, 40, 64, Big Bore, iCT and iCT SP are Computed Tomography X-Ray Systems. The devices are Computed Tomography Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes
Reason for recall
Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.