Recall Z-2032-2025— Baxter Healthcare Corporation

Class I · Ongoing

Class I FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-05-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.. Reason: The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop..

Recalling firm
Baxter Healthcare Corporation
Classification
Class I
Status
Ongoing
Initiated
2025-05-30
Firm location
Deerfield, IL, United States

Product description

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Data sourced from openFDA. This site is unofficial and independent of the FDA.