Recall Z-2032-2025— Baxter Healthcare Corporation
Class I · Ongoing
Class I FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-05-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.. Reason: The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-05-30
- Firm location
- Deerfield, IL, United States
Product description
Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
Reason for recall
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.