Recall Z-2026-2014— Steris Corporation

Class II · Terminated

Class II FDA medical device recall by Steris Corporation, initiated 2014-06-27, terminated 2015-09-16. FDA's recall record names no specific 510(k) clearance for this event. Product: The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.. Reason: STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked..

Recalling firm
Steris Corporation
Classification
Class II
Status
Terminated
Initiated
2014-06-27
Terminated
2015-09-16
Firm location
Mentor, OH, United States

Product description

The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.

Reason for recall

STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.