Recall Z-2016-2021— Boston Scientific Corporation
Class I · Ongoing
Class I FDA medical device recall by Boston Scientific Corporation, initiated 2021-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021). Reason: The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population.
- Recalling firm
- Boston Scientific Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2021-06-03
- Firm location
- Saint Paul, MN, United States
Product description
INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021)
Reason for recall
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
Data sourced from openFDA. This site is unofficial and independent of the FDA.