Recall Z-2015-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2026-03-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MAJOR NEURO PACK-LF, Medline Kit Number/SKU DYNJ0578916X; 2. ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ51637D; 3. NEURO SEO, Medline Kit Number/SKU DYNJ65650.. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2026-03-19
Firm location
Northfield, IL, United States

Product description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MAJOR NEURO PACK-LF, Medline Kit Number/SKU DYNJ0578916X; 2. ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ51637D; 3. NEURO SEO, Medline Kit Number/SKU DYNJ65650.

Reason for recall

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.