Recall Z-2015-2021— Boston Scientific Corporation

Class I · Ongoing

Class I FDA medical device recall by Boston Scientific Corporation, initiated 2021-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: ADVANTIO DR EL (pacemaker), Model Numbers: a) J064 b) J067 c) K064 (added 7/15/2021) d) K084 (added 7/15/2021) e) K087 (added 7/15/2021). Reason: The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population.

Recalling firm
Boston Scientific Corporation
Classification
Class I
Status
Ongoing
Initiated
2021-06-03
Firm location
Saint Paul, MN, United States

Product description

ADVANTIO DR EL (pacemaker), Model Numbers: a) J064 b) J067 c) K064 (added 7/15/2021) d) K084 (added 7/15/2021) e) K087 (added 7/15/2021)

Reason for recall

The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population

Data sourced from openFDA. This site is unofficial and independent of the FDA.