Recall Z-2013-2019— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-06-12, terminated 2020-09-18. It names one FDA 510(k) clearance: K052202. Product: AXIOM Artis FA, Model Number 5904441. Reason: A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-12
- Terminated
- 2020-09-18
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
AXIOM Artis FA, Model Number 5904441
Reason for recall
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Data sourced from openFDA. This site is unofficial and independent of the FDA.