Recall Z-2011-2024— Atrium Medical Corporation

Class II · Ongoing

Class II FDA medical device recall by Atrium Medical Corporation, initiated 2024-05-03. It names 2 FDA 510(k) clearances: K071923, K060124. Product: FLIXENE, 6X30, 1GDS, STR. Single-Ended Slider GDS. Vascular graft. Reason: Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces..

Recalling firm
Atrium Medical Corporation
Classification
Class II
Status
Ongoing
Initiated
2024-05-03
Firm location
Merrimack, NH, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FLIXENE, 6X30, 1GDS, STR. Single-Ended Slider GDS. Vascular graft

Reason for recall

Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.