Recall Z-2007-2023— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2023-05-19. FDA's recall record names no specific 510(k) clearance for this event. Product: NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed. Reason: Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-05-19
- Firm location
- Deerfield, IL, United States
Product description
NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed
Reason for recall
Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.
Data sourced from openFDA. This site is unofficial and independent of the FDA.