Recall Z-2000-2021— Boston Scientific Corporation

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2021-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331. Reason: Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators..

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Ongoing
Initiated
2021-06-03
Firm location
Saint Paul, MN, United States

Product description

Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331

Reason for recall

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.