Recall Z-1998-2026— Heraeus Medical GmbH (Dental Division)

Class III · Ongoing

Class III FDA medical device recall by Heraeus Medical GmbH (Dental Division), initiated 2026-03-25. FDA's recall record names no specific 510(k) clearance for this event. Product: Heraeus, PALAMIX duo. Material Number: 66057897.. Reason: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months)..

Recalling firm
Heraeus Medical GmbH (Dental Division)
Classification
Class III
Status
Ongoing
Initiated
2026-03-25
Firm location
Wehrheim, Germany

Product description

Heraeus, PALAMIX duo. Material Number: 66057897.

Reason for recall

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).