Recall Z-1997-2019— Philips Medical Systems Nederlands

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems Nederlands, initiated 2018-11-30, terminated 2024-03-27. It names one FDA 510(k) clearance: K172822. Product: Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2. Reason: Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images..

Recalling firm
Philips Medical Systems Nederlands
Classification
Class II
Status
Terminated
Initiated
2018-11-30
Terminated
2024-03-27
Firm location
Best, N/A, Netherlands

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2

Reason for recall

Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.

Data sourced from openFDA. This site is unofficial and independent of the FDA.