Recall Z-1996-2021— Flowonix Medical Inc
Class II · Terminated
Class II FDA medical device recall by Flowonix Medical Inc, initiated 2021-05-26, terminated 2024-02-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242. Reason: Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection.
- Recalling firm
- Flowonix Medical Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-05-26
- Terminated
- 2024-02-20
- Firm location
- Mansfield, MA, United States
Product description
Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
Reason for recall
Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
Data sourced from openFDA. This site is unofficial and independent of the FDA.