Recall Z-1994-2025— Barco N.V.
Class II · Ongoing
Class II FDA medical device recall by Barco N.V., initiated 2025-05-09. It names one FDA 510(k) clearance: K173381. Product: Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system. Reason: Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures.
- Recalling firm
- Barco N.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-05-09
- Firm location
- Kortrijk, Belgium
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
Reason for recall
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
Data sourced from openFDA. This site is unofficial and independent of the FDA.