Recall Z-1993-2016— Physio-Control, Inc.

Class II · Terminated

Class II FDA medical device recall by Physio-Control, Inc., initiated 2016-01-27, terminated 2017-03-31. It names one FDA 510(k) clearance: K142430. Product: LIFEPAK15 monitor/defibrillator with End-Tidal C02 (EtC02) feature installed. Part numbers: V15-2-xxxxxx (includes software version 1 and version 2); V15-5-xxxxxx (inclues software version 4). The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.. Reason: The firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a portion of the leading digit of the EtC02 value. This affects all LIFEPAK 15 with an EtC02 feature installed and configured to the kPa or % setting..

Recalling firm
Physio-Control, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-01-27
Terminated
2017-03-31
Firm location
Redmond, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

LIFEPAK15 monitor/defibrillator with End-Tidal C02 (EtC02) feature installed. Part numbers: V15-2-xxxxxx (includes software version 1 and version 2); V15-5-xxxxxx (inclues software version 4). The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.

Reason for recall

The firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a portion of the leading digit of the EtC02 value. This affects all LIFEPAK 15 with an EtC02 feature installed and configured to the kPa or % setting.