Recall Z-1990-2021— Philips North America Llc

Class II · Ongoing

Class II FDA medical device recall by Philips North America Llc, initiated 2021-05-01. It names one FDA 510(k) clearance: K060937. Product: Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728306 - Brilliance iCT 728311 - Brilliance iCT SP. Reason: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury..

Recalling firm
Philips North America Llc
Classification
Class II
Status
Ongoing
Initiated
2021-05-01
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728306 - Brilliance iCT 728311 - Brilliance iCT SP

Reason for recall

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.