Recall Z-1979-2019— Roche Diagnostics Corporation
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2019-06-03, terminated 2020-08-28. It names one FDA 510(k) clearance: K040435. Product: The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.. Reason: Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent..
- Recalling firm
- Roche Diagnostics Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-03
- Terminated
- 2020-08-28
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Reason for recall
Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.