Recall Z-1979-2013— Boston Scientific CRM Corp

Class II · Terminated

Class II FDA medical device recall by Boston Scientific CRM Corp, initiated 2013-07-15, terminated 2013-11-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Boston Scientific, ENERGEN DR ICD, Model E143. The device is an Implantable Cardioverter Defibrillator.. Reason: Boston Scientific CRM manufacturing quality system recently discovered test artifacts (related to shock charge time) in a small number of implantable defibrillators (ENERGEN DR ICD model E143) which require further investigation. While these devices successfully passed all manufacturing tests and met all requirements, they showed a slightly longer charge time than other devices during a manufactu.

Recalling firm
Boston Scientific CRM Corp
Classification
Class II
Status
Terminated
Initiated
2013-07-15
Terminated
2013-11-12
Firm location
Saint Paul, MN, United States

Product description

Boston Scientific, ENERGEN DR ICD, Model E143. The device is an Implantable Cardioverter Defibrillator.

Reason for recall

Boston Scientific CRM manufacturing quality system recently discovered test artifacts (related to shock charge time) in a small number of implantable defibrillators (ENERGEN DR ICD model E143) which require further investigation. While these devices successfully passed all manufacturing tests and met all requirements, they showed a slightly longer charge time than other devices during a manufactu

Data sourced from openFDA. This site is unofficial and independent of the FDA.