Recall Z-1976-2019— Cardiocommand Inc.
Class III · Terminated
Class III FDA medical device recall by Cardiocommand Inc., initiated 2019-06-24, terminated 2021-04-01. FDA's recall record names no specific 510(k) clearance for this event. Product: TAPSYSTEM Model 2A, REF 8002A. Reason: There is a potential for an assembly error resulting in the battery wire being connected to the ON/OFF power switch incorrectly..
- Recalling firm
- Cardiocommand Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2019-06-24
- Terminated
- 2021-04-01
- Firm location
- Tampa, FL, United States
Product description
TAPSYSTEM Model 2A, REF 8002A
Reason for recall
There is a potential for an assembly error resulting in the battery wire being connected to the ON/OFF power switch incorrectly.
Data sourced from openFDA. This site is unofficial and independent of the FDA.