Recall Z-1970-2023— LumiraDx

Class II · Ongoing

Class II FDA medical device recall by LumiraDx, initiated 2023-04-25. FDA's recall record names no specific 510(k) clearance for this event. Product: SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB. Reason: Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab.

Recalling firm
LumiraDx
Classification
Class II
Status
Ongoing
Initiated
2023-04-25
Firm location
Waltham, MA, United States

Product description

SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB

Reason for recall

Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab