Recall Z-1969-2026— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2026-04-14. It names 4 FDA 510(k) clearances: K251397, K223458, K213583 and 1 more. Product: Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.. Reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small..
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-04-14
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
4 on FDA's record- K251397 — Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR 5300 and MR 7700 MR System
- K223458 — Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
- K213583 — Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR Systems
- K213516 — Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
Reason for recall
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Data sourced from openFDA. This site is unofficial and independent of the FDA.