Recall Z-1969-2013— Terumo Cardiovascular Systems Corporation

Class II · Terminated

Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2013-07-26, terminated 2017-02-07. It names one FDA 510(k) clearance: K022947. Product: Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.. Reason: The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react..

Recalling firm
Terumo Cardiovascular Systems Corporation
Classification
Class II
Status
Terminated
Initiated
2013-07-26
Terminated
2017-02-07
Firm location
Ann Arbor, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.

Reason for recall

The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

Data sourced from openFDA. This site is unofficial and independent of the FDA.