Recall Z-1966-2012— Terumo Cardiovascular Systems Corporation

Class II · Terminated

Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2012-06-22, terminated 2013-11-08. It names one FDA 510(k) clearance: K022947. Product: System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The System 1 base includes an integrated backup battery in the event of power loss.. Reason: Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage..

Recalling firm
Terumo Cardiovascular Systems Corporation
Classification
Class II
Status
Terminated
Initiated
2012-06-22
Terminated
2013-11-08
Firm location
Ann Arbor, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The System 1 base includes an integrated backup battery in the event of power loss.

Reason for recall

Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage.