Recall Z-1963-2014— Baxter Healthcare Corp.

Class II · Terminated

Class II FDA medical device recall by Baxter Healthcare Corp., initiated 2014-06-10, terminated 2017-02-22. FDA's recall record names no specific 510(k) clearance for this event. Product: EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.. Reason: If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag..

Recalling firm
Baxter Healthcare Corp.
Classification
Class II
Status
Terminated
Initiated
2014-06-10
Terminated
2017-02-22
Firm location
Deerfield, IL, United States

Product description

EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.

Reason for recall

If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.

Data sourced from openFDA. This site is unofficial and independent of the FDA.