Recall Z-1961-2014— Integra LifeSciences Corp.
Class II · Terminated
Class II FDA medical device recall by Integra LifeSciences Corp., initiated 2014-06-09, terminated 2014-11-06. It names one FDA 510(k) clearance: K083089. Product: MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws. Reason: Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique..
- Recalling firm
- Integra LifeSciences Corp.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-06-09
- Terminated
- 2014-11-06
- Firm location
- Plainsboro, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws
Reason for recall
Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.