Recall Z-1956-2021— Philips Respironics, Inc.
Class I · Ongoing
Class I FDA medical device recall by Philips Respironics, Inc., initiated 2021-06-14. It names 3 FDA 510(k) clearances: K111610, K093416, K083526. Product: Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.. Reason: The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals..
- Recalling firm
- Philips Respironics, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2021-06-14
- Firm location
- Murrysville, PA, United States
510(k) clearances named in this recall
3 on FDA's record- K111610 — TRILOGY SERIES VENTILATOR WITH OXIMETRY
- K093416 — TRILOGY 200 VENTILATOR
- K083526 — TRILOGY 100 VENTILATOR
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
Reason for recall
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.