Recall Z-1955-2012— Civco Medical Instruments Inc
Class II · Terminated
Class II FDA medical device recall by Civco Medical Instruments Inc, initiated 2012-05-07, terminated 2015-01-07. It names one FDA 510(k) clearance: K092619. Product: CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor, for use with 3D Guidance medSAFE trackers, Ref. 610-1067. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices. Reason: The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device..
- Recalling firm
- Civco Medical Instruments Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-05-07
- Terminated
- 2015-01-07
- Firm location
- Kalona, IA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor, for use with 3D Guidance medSAFE trackers, Ref. 610-1067. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices
Reason for recall
The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.
Data sourced from openFDA. This site is unofficial and independent of the FDA.