Recall Z-1951-2024— Atrium Medical Corporation

Class II · Ongoing

Class II FDA medical device recall by Atrium Medical Corporation, initiated 2024-05-03. It names one FDA 510(k) clearance: K992958. Product: ADVANTA VXT, 5X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft. Reason: Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces..

Recalling firm
Atrium Medical Corporation
Classification
Class II
Status
Ongoing
Initiated
2024-05-03
Firm location
Merrimack, NH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVANTA VXT, 5X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

Reason for recall

Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

Data sourced from openFDA. This site is unofficial and independent of the FDA.