Recall Z-1948-2014— Ventlab LLC
Class I · Terminated
Class I FDA medical device recall by Ventlab LLC, initiated 2014-05-16, terminated 2015-11-17. It names 3 FDA 510(k) clearances: K012842, K982215, K953546. Product: Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.. Reason: The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim..
- Recalling firm
- Ventlab LLC
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2014-05-16
- Terminated
- 2015-11-17
- Firm location
- Grand Rapids, MI, United States
510(k) clearances named in this recall
3 on FDA's record- K012842 — V*CARE MANUAL RESUSCITATOR
- K982215 — MANUAL RESUSCITATOR WITH ACE
- K953546 — MANUAL PULMONARY RESUSICITATOR
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.
Reason for recall
The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.