Recall Z-1947-2024— LEICA BIOSYSTEMS NUSSLOCH GMBH
Class II · Ongoing
Class II FDA medical device recall by LEICA BIOSYSTEMS NUSSLOCH GMBH, initiated 2024-04-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.. Reason: Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks..
- Recalling firm
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-04-10
- Firm location
- Nussloch, N/A, Germany
Product description
Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.
Reason for recall
Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.