Recall Z-1946-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-11-27, terminated 2019-04-19. It names one FDA 510(k) clearance: K913123. Product: Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.. Reason: Lot J1 of the Syva¿ EMIT¿ 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-11-27
Terminated
2019-04-19
Firm location
Newark, DE, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.

Reason for recall

Lot J1 of the Syva¿ EMIT¿ 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.