Recall Z-1944-2018— Olsen Medical

Class II · Terminated

Class II FDA medical device recall by Olsen Medical, initiated 2018-05-11, terminated 2019-04-15. It names one FDA 510(k) clearance: K130669. Product: Symmetry Olsen bipolar forceps: (a1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311, QTY 5 (b1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311R, QTY 5 (c1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.0mm Tips, REF 20-0321, QTY 5 (d1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.5mm Tips, REF 20-0331, QTY 5 (e1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth, 1.5mm Tips, REF 20-0331R, QTY 5 (f1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth Micro Tips, REF 20-0351, QTY 5 (g1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 0.5mm Tips, REF 20-0361, QTY 5 (h1) Insulated Irrigating Bayonet Forceps 7-1/8" (181.0mm) 0.5mm Tips, REF 20-0361G, QTY 5 (i1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 0.5mm Tips, REF 20-0361R, QTY 5 (j1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.0mm Tips, REF 20-0371, QTY 5 (k1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.0mm Tips, REF 20-0371R, QTY 5 (l1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.5mm Tips, REF 20-0381, QTY 5 (m1)Insulated Irrigating Bayonet Forceps 7-1/8" (181.0mm) 1.5mm Tips, REF 20-0381G, QTY 5 (n1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.5mm Tips, REF 20-0381R, QTY 5 (o1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 0.5mm Tips, REF 20-0411, QTY 5 (p1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 0.5mm Tips, REF 20-0411R, QTY 5 (q1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.0mm Tips, REF 20-0421, QTY 5 (r1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.0mm Tips, REF 20-0421R, QTY 5 (s1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.5mm Tips, REF 20-0431, QTY 5 (t1) Insulated Irrigating Bayonet Forceps 8.25" (309.6mm) 1.5mm Tips, REF 20-0431G, QTY 5 (u1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.5mm Tips, REF 20-0431R, QTY 5 (v1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.0mm Tips, REF 20-0481, QTY 5 (w1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.0mm Tips, REF 20-0481R, QTY 5 (x1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.5mm Tips, REF 20-0491, QTY 5 (y1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.5mm Tips, REF 20-0491R (z1) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 0.5mm Tips, REF 20-0571IMT, QTY 5 (a2) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 1.0mm Tips, REF 20-0572IMT, QTY 5 (b2) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 1.5mm Tips, REF 20-0573IMT, QTY 5 (c2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 0.5mm Tips, REF 20-0581IMT, QTY 5 (d2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 1.0mm Tips, REF 20-0582IMT, QTY 5 (e2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 1.5mm Tips, REF 20-0583IMT, QTY 5 (f2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 0.5mm Tips, REF 20-0591IMT, QTY 5 (g2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 1.0mm Tips, REF 20-0592IMT, QTY 5 (h2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 1.5mm Tips, REF 20-0593IMT (i2) Insulated Coaptation Forceps 4.5" (11.4cm) Smooth Micro Tips, Angled, REF 20-1010KI, QTY 10 (j2) Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, REF 20-1030K, QTY 10 (k2) Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, Angled 45o, REF 20-1040K, QTY 10 (l2) Insulated Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, Angled 45o, REF 20-1040KI, QTY 10 (m2) Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060K, QTY 10 (n2) Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060K-28, QTY 10 (o2) Insulated Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060KI, QTY 10 (p2) Insulated Jewelers Forceps 4.5" (11.4cm) Smooth Micro Ti. Reason: Products may encroach into the sterile pouch seal thereby increasing the risk of the seal being compromised..

Recalling firm
Olsen Medical
Classification
Class II
Status
Terminated
Initiated
2018-05-11
Terminated
2019-04-15
Firm location
Louisville, KY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Symmetry Olsen bipolar forceps: (a1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311, QTY 5 (b1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311R, QTY 5 (c1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.0mm Tips, REF 20-0321, QTY 5 (d1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.5mm Tips, REF 20-0331, QTY 5 (e1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth, 1.5mm Tips, REF 20-0331R, QTY 5 (f1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth Micro Tips, REF 20-0351, QTY 5 (g1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 0.5mm Tips, REF 20-0361, QTY 5 (h1) Insulated Irrigating Bayonet Forceps 7-1/8" (181.0mm) 0.5mm Tips, REF 20-0361G, QTY 5 (i1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 0.5mm Tips, REF 20-0361R, QTY 5 (j1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.0mm Tips, REF 20-0371, QTY 5 (k1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.0mm Tips, REF 20-0371R, QTY 5 (l1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.5mm Tips, REF 20-0381, QTY 5 (m1)Insulated Irrigating Bayonet Forceps 7-1/8" (181.0mm) 1.5mm Tips, REF 20-0381G, QTY 5 (n1) Insulated Irrigating Bayonet Forceps 7-1/8" (18.1cm) Smooth 1.5mm Tips, REF 20-0381R, QTY 5 (o1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 0.5mm Tips, REF 20-0411, QTY 5 (p1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 0.5mm Tips, REF 20-0411R, QTY 5 (q1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.0mm Tips, REF 20-0421, QTY 5 (r1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.0mm Tips, REF 20-0421R, QTY 5 (s1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.5mm Tips, REF 20-0431, QTY 5 (t1) Insulated Irrigating Bayonet Forceps 8.25" (309.6mm) 1.5mm Tips, REF 20-0431G, QTY 5 (u1) Insulated Irrigating Bayonet Forceps 8.25" (21.0cm) Smooth 1.5mm Tips, REF 20-0431R, QTY 5 (v1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.0mm Tips, REF 20-0481, QTY 5 (w1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.0mm Tips, REF 20-0481R, QTY 5 (x1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.5mm Tips, REF 20-0491, QTY 5 (y1) Insulated Irrigating Bayonet Forceps 10.5" (26.7cm) Smooth 1.5mm Tips, REF 20-0491R (z1) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 0.5mm Tips, REF 20-0571IMT, QTY 5 (a2) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 1.0mm Tips, REF 20-0572IMT, QTY 5 (b2) Precision Non-Stick Insulated Irrigating Forceps 7" (17.8cm) 1.5mm Tips, REF 20-0573IMT, QTY 5 (c2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 0.5mm Tips, REF 20-0581IMT, QTY 5 (d2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 1.0mm Tips, REF 20-0582IMT, QTY 5 (e2) Precision Non-Stick Insulated Irrigating Forceps 8" (20.3cm) 1.5mm Tips, REF 20-0583IMT, QTY 5 (f2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 0.5mm Tips, REF 20-0591IMT, QTY 5 (g2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 1.0mm Tips, REF 20-0592IMT, QTY 5 (h2) Precision Non-Stick Insulated Irrigating Forceps 9" (22.9cm) 1.5mm Tips, REF 20-0593IMT (i2) Insulated Coaptation Forceps 4.5" (11.4cm) Smooth Micro Tips, Angled, REF 20-1010KI, QTY 10 (j2) Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, REF 20-1030K, QTY 10 (k2) Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, Angled 45o, REF 20-1040K, QTY 10 (l2) Insulated Iris Forceps 4.5" (11.4cm) Serrated 1.0mm Tips, Angled 45o, REF 20-1040KI, QTY 10 (m2) Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060K, QTY 10 (n2) Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060K-28, QTY 10 (o2) Insulated Jewelers Forceps 4.5" (11.4cm) Smooth Micro Tips, REF 20-1060KI, QTY 10 (p2) Insulated Jewelers Forceps 4.5" (11.4cm) Smooth Micro Ti

Reason for recall

Products may encroach into the sterile pouch seal thereby increasing the risk of the seal being compromised.