Recall Z-1940-2024— Stryker Neurovascular
Class II · Ongoing
Class II FDA medical device recall by Stryker Neurovascular, initiated 2024-04-18. It names one FDA 510(k) clearance: K202522. Product: The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.. Reason: Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique..
- Recalling firm
- Stryker Neurovascular
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-04-18
- Firm location
- Fremont, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.
Reason for recall
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
Data sourced from openFDA. This site is unofficial and independent of the FDA.