Recall Z-1936-2025— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2025-05-29. FDA's recall record names no specific 510(k) clearance for this event. Product: SmartPath to dStream for 1.5T¿; Product Number: 782146;. Reason: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage..

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2025-05-29
Firm location
Cambridge, MA, United States

Product description

SmartPath to dStream for 1.5T¿; Product Number: 782146;

Reason for recall

identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.