Recall Z-1935-2013— Synthes USA HQ, Inc.

Class II · Terminated

Class II FDA medical device recall by Synthes USA HQ, Inc., initiated 2013-06-05, terminated 2015-09-21. It names one FDA 510(k) clearance: K020517. Product: 5.0mm Unit Rod 270mm 5.0mm Unit Rod 290mm 5.0mm Unit Rod 310mm 5.0mm Unit Rod 330mm 5.0mm Unit Rod 350mm 5.0mm Unit Rod 370mm 5.0mm Unit Rod 390mm 5.0mm Unit Rod 410mm 5.0mm Unit Rod 430mm 5.0mm Unit Rod 450mm pedicle screw spinal system; Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.. Reason: DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product was produced using a finishing process not identified as part of the manufacturing specification. No adverse event reports associated with the use of this product were received, and an extensive engineering and medical safety review concluded that the product poses no incremental risk to t.

Recalling firm
Synthes USA HQ, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-06-05
Terminated
2015-09-21
Firm location
West Chester, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

5.0mm Unit Rod 270mm 5.0mm Unit Rod 290mm 5.0mm Unit Rod 310mm 5.0mm Unit Rod 330mm 5.0mm Unit Rod 350mm 5.0mm Unit Rod 370mm 5.0mm Unit Rod 390mm 5.0mm Unit Rod 410mm 5.0mm Unit Rod 430mm 5.0mm Unit Rod 450mm pedicle screw spinal system; Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Reason for recall

DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product was produced using a finishing process not identified as part of the manufacturing specification. No adverse event reports associated with the use of this product were received, and an extensive engineering and medical safety review concluded that the product poses no incremental risk to t

Data sourced from openFDA. This site is unofficial and independent of the FDA.