Recall Z-1933-2016— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2016-04-13, terminated 2016-06-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Cholesterol, Catalog No. OSR6x16 Product Usage:Cholesterol OSR6x16 is intended as a System reagent for the quantitative determination of Cholesterol concentrations in human serum on Beckman Coulter AU analyzers.. Reason: Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-04-13
- Terminated
- 2016-06-30
- Firm location
- Brea, CA, United States
Product description
Cholesterol, Catalog No. OSR6x16 Product Usage:Cholesterol OSR6x16 is intended as a System reagent for the quantitative determination of Cholesterol concentrations in human serum on Beckman Coulter AU analyzers.
Reason for recall
Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.