Recall Z-1932-2016— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2016-04-13, terminated 2016-06-30. It names one FDA 510(k) clearance: K062862. Product: Uric Acid, Catalog No. OSR6x98 Product Usage:Uric acid OSR6x98 is intended as a System reagent for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on Beckman Coulter AU analyzers.. Reason: Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-04-13
- Terminated
- 2016-06-30
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Uric Acid, Catalog No. OSR6x98 Product Usage:Uric acid OSR6x98 is intended as a System reagent for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on Beckman Coulter AU analyzers.
Reason for recall
Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.
Data sourced from openFDA. This site is unofficial and independent of the FDA.