Recall Z-1931-2018— Medline Industries Inc

Class II · Terminated

Class II FDA medical device recall by Medline Industries Inc, initiated 2018-04-24, terminated 2020-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE Sterile* Needle Counter, 100Ct. Double Magnet / Foam Block, reorder: DYNJNC100F Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.. Reason: The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters..

Recalling firm
Medline Industries Inc
Classification
Class II
Status
Terminated
Initiated
2018-04-24
Terminated
2020-08-04
Firm location
Northfield, IL, United States

Product description

MEDLINE Sterile* Needle Counter, 100Ct. Double Magnet / Foam Block, reorder: DYNJNC100F Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.

Reason for recall

The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.

Data sourced from openFDA. This site is unofficial and independent of the FDA.